Product Number: XD-(5-30)VOC-AHF/AF
Brand Name:MODERNER
Product Origin: SHANGHAI
Purchasing Heat:2598
Standard:
The climate chamber is started with one key and runs fully automatically. Temperature and relative humidity control adopts PLC as the core control unit, with high reliability and strong anti-interference

The walk-in VOC emission detection environmental test chamber is to restore the parameters such as temperature, humidity, ventilation times, and air flow rate on the surface of the sample through artificial simulation, and test the actual situation through the test methods and verification methods specified in the standard. The VOC release amount of the sample in the application is analyzed to determine whether it meets the corresponding environmental protection evaluation standards, and the release amount of volatile toxic and harmful (VOC) substances contained in the sample material and the impact of its release characteristics on the environment can be measured.
The climatic chamber is set up with a one-touch start mode for fully automatic operation. The temperature, relative humidity and replacement air flow are controlled by a PLC as the core control unit, which is highly reliable and resistant to interference.
Real-time monitoring and data storage of all parameters during operation, historical data and curves for each point of the test can be checked and viewed in real time, and can also be downloaded and saved to a PC for data analysis, editing and archiving via a mobile storage device.
XD-5VOC-AHF 5 cubic Large walk-in Environmental Chamber



| Type | XD-(5-30)VOC-AHF |
| Environmental cabin volume | 5-30m³ (Size can be customized according to user's requirements) |
| Temperature adjustable range | 20 ~ 65℃ |
| Humidity adjustable range | 30~80%RH(20-30℃) |
| Temperature fluctuation | ≤±0.5℃ |
| Temperature deviation | ≤±0.5℃ |
| Humidity fluctuation | ≤±3%RH |
| Humidity deviation | ≤±3%RH |
| Air replacement rate | 1~3 Times/hour (adjustable) |
| Cabin pressure | 10±5Pa |
| Airtightness | At 1kPa (gauge pressure), the cabin air leakage rate VL is less than or equal to 0.5 %× cabin capacity /min, or the cabin air leakage rate VL is less than or equal to 5 %× air supply rate VS. |
| Stabilization time | The temperature stabilization time is less than 40min, and the humidity stabilization time is less than 60min. The equipment can run continuously for not less than 30 days; |
| Background concentration | Under no load, the background concentration of formaldehyde is ≤ 0.006 mg/m³, TVOC is ≤ 0.002 mg/m³, and the background concentration of monomer VOC is ≤ 0.002 mg/m³. |
| Noise | ≤60 dB (A sound level) |
| External environmental requirements | -10℃~ 40℃;Relative humidity:20-90% |
| Type | XD-(5-30)VOC-AHF |
| Machine | 1 |
| Electrical schematic diagram | 1 |
| Instructions | 1 |
| Certificate | 1 |
| Type | Name |
| XD-(5-30)VOC-AHF | Large walk-in Environmental Chamber |