Product Number: XD-5RVOC-AHF/AF
Brand Name:MODERNER
Product Origin: SHANGHAI
Purchasing Heat:1200
Standard:
The climate chamber is started with one key and runs fully automatically. Temperature and relative humidity control adopts PLC as the core control unit, with high reliability and strong anti-interference
The walk-in VOC emission detection environmental test chamber is to restore the parameters such as temperature, humidity, ventilation times, and air flow rate on the surface of the sample through artificial simulation, and test the actual situation through the test methods and verification methods specified in the standard. The VOC release amount of the sample in the application is analyzed to determine whether it meets the corresponding environmental protection evaluation standards, and the release amount of volatile toxic and harmful (VOC) substances contained in the sample material and the impact of its release characteristics on the environment can be measured.
The climatic chamber is set up with a one-touch start mode for fully automatic operation. The temperature, relative humidity and replacement air flow are controlled by a PLC as the core control unit, which is highly reliable and resistant to interference.
Real-time monitoring and data storage of all parameters during operation, historical data and curves for each point of the test can be checked and viewed in real time, and can also be downloaded and saved to a PC for data analysis, editing and archiving via a mobile storage device.
XD-5RVOC-AHF 5 cubic Large walk-in Environmental Chamber
Type | XD-5RVOC-AHF |
Studio size W*L*H | Inner box size5±0.02 m3;1600*1600*2000mm |
Dimensions H*D*W | 2280*2700mm*1800mm |
Temperature/Humidity | temperature range:20℃~65℃; Humidity range:30%~75%RH(at +20~+60℃) |
Meter resolution | temperature :0.1℃;Humidity:0.1%RH; |
Temperature Fluctuation | ±0.2℃; |
Temperature Uniformity | ≤1.5℃ |
Humidity Fluctuation | ≤±1.5%RH |
Air replacement rate | 1~3 times/hour (adjustable) |
Flowmeter Accuracy | Digital flowmeter, accuracy class 0.5; 4-40m3/L |
Air Tightness | Ø When the overpressure is 1000pa, the leakage is less than 10-3×1m³/min; Ø The constant pressure in the cabin is 10pa ±5pa Ø The gas flow difference between inlet and outlet gas is less than 1%; Ø Air pressure monitoring in the box, analog digital micro pressure sensor, touch screen online pressure display box in the box |
Stable Schedule | Temperature stabilization time is less than 40min, humidity stabilization time is less than 60min, |
Background Indicator | TVOC≤0.02 mg/m³ |
Noise | ≤60dB (A sound level) |
Factory settings | Cabin temperature/humidity 23°C/50% RH, air exchange rate 1 time/h, cabin air pressure 10 Pa |
5 cubic air high temperature purification device
Type | XD-5KQ-HT02 |
Air Flow | >15m³/h |
Air Supply Pressure | 0.1~0.24bar |
Supply Air Temperature | RT |
Formaldehyde Mass Concentration | ≤±0.001mg/m³ |
VOC Concentration | TVOC≤0.01mg/m³,VOC≤0.001mg/m³ |
Noise | ≤60db (A sound level) |
Type | XD-5RVOC-AHF |
Machine | 1 |
Electrical schematic diagram | 1 |
Instructions | 1 |
certificate | 1 |
Type | Name |
XD-5RVOC-AHF | Large walk-in Environmental Chamber |
XD-5KQ-HT02 | 5 Cubic Air High Temperature Turification Device |
Factory settings | Incineration temperature 350°C, gas supply temperature RT room temperature |