XDWR-1-3FATemperature resistance test chamber

Temperature resistance test chamber

Product Number: XDWR-1-3FA

Brand Name:MODERNER

Product Origin: China

Purchasing Heat:1501

Standard:

The equipment has three temperature control areas, namely constant temperature water control box, low temperature control box, high temperature control box.

Click to see the big picture

Product Details

 Ⅰ.Introduction:

The equipment has three temperature control areas, namely constant temperature water control box, low temperature control box, high temperature control box. The equipment basket will put the samples into the three environmental compartments respectively, and the circulation meter will be used to evaluate whether the coating of the test plate has the phenomena of pulverization, cracking, peeling, foaming, discoloration, etc. 


 


 


 


 


 

Ⅱ. Main Technical Parameters: 

Model

XDWR-1-3FA

Installation Dimensions

2600×1100×1860mm 

Hanging basket liner size *3

370×280×260mm(W×D×H)

Instrument  resolution

Temperature:0.1℃

Temperature Range

1、Constant temperature water:20~30℃

2、Low temperature box:-25~20℃

3、High temperature box:At room temperature~70℃

Temperature fluctuation

±1℃

Temperature uniformity

≤2℃

Air displacement rate

0.5~3 t/h(adjustable)

Stable time

Temperature stabilization time<40min

Noise

≤75db,

Box door and equipment door will not have resonance sound

Net weight

600KG

 

Equipment requirements

Electrical source

① Power Source  : AC380V   50Hz

② Maximum load capacity : 10KW

③  The biggest electricity flow  : 30A

④ Power switching capacity : 32A

To the water

Please use pure water or distilled water.   Electrical conductivity≤10μs/cm²    

Move in condition

The components of the laboratory,

When the department moves in,

Please satisfy the conditions on the right

The height of the passageway and door:More than 1200 mm

The width of the passageway and door:More than 1900 mm

When using the elevator; In addition to the above dimensions,

the following requirements must be met

Depth: more than 2300mm; Loading weight: more than 1000Kg

※ If the above conditions cannot be met, please be sure to contact the company in advance.

 

Ⅲ . Configuration list:

Type

XDWR-1-3FA

Machine

1

Electrical schematic diagram

1

Instructions

1

Certificate

1

 

Ⅳ. Ordering Information:

Model

Name

XDWR-1-3FA

Temperature resistance test chamber

 