XD-1VOC-A11m3 Emission Test Chamber

1m3 Emission Test Chamber

Product Number: XD-1VOC-A1

Brand Name:MODERNER

Product Origin: SHANGHAI

Purchasing Heat:438

Standard: ASTM D 5116-97,ASTM D 6007-02,ASTM D 6330-98,ASTM D 6670-01,ASTM D 6803-02 ,EN 717-1 2004,ENV 13419-1 ,ISO 12460-1.2007 ,ISO 16000-9-2006,ANSI/BIFMAX7.1-2007

Test Chamber can simulate real environment parameters, such as temperature, relative humidity, air change rate, air flow, etc. Through analyzing test data, it can measure emission rate of volatile organic compounds (VOC) and its release characteristics impact on the environment.

Click to see the big picture

Product Details

Ⅰ.Function overview:

Test Chamber can simulate real environment parameters, such as temperature, relative humidity, air change rate, air flow, etc. Through analyzing test data, it can measure emission rate of volatile organic compounds (VOC) and its release characteristics impact on the environment, this can: 

  • Determine the effects of environmental variables on emission rate and release characteristics of VOC.
  • The grades of materials are determined according to the VOC emission characteristics of the sample materials.
  • R & D of sample materials and verification of indoor air pollution characteristics from VOC emission. 
  • Emission rate of VOC in sample experiences time decay in natural environment. 
  • Assessment of indoor air quality is predicted based on test data analysis.


 


 


 

  
 

II.Meets standards:

ASTM D 5116-97
Standard Guide for Small-Scale Environmental Chamber Determinations of Organic Emissions from Indoor Materials/Products (Withdrawn)
ASTM D 6007-02
 Standard Test Method for Determining Formaldehyde Concentration in Air from Wood Products Using a Small Scale Chamber
ASTM D 6330-98
Standard Practice for Determination of Volatile Organic Compounds (Excluding Formaldehyde) Emissions from Wood-Based Panels Using  Small Environmental Chambers Under Defined Test Conditions
ASTM D 6670-01
 Standard Practice for Full-Scale Chamber Determination of Volatile Organic Emissions from Indoor Materials/Products
ASTM D 6803-02
Standard Practice for Testing and Sampling of Volatile Organic Compounds (Including Carbonyl Compounds) Emitted from Paint Using  Small Environmental Chambers
EN 717-1 2004
Wood-Based Panels - Determination Of Formaldehyde Release - Part 1: Formaldehyde Emission By The Chamber Method
ENV 13419-1 
Determinationoftheemissionofvolatileorganiccompounds-Part1:Emissiontestchambermethod
ISO 12460-1.2007
specifies a 1 m3 chamber method for the determination of the formaldehyde emission from wood-based panels under defined conditions,  relating to typical conditions in real-life.
ISO 16000-9-2006
Indoor air -Part 9: Determination of the emission of volatile organic compounds from building products and furnishing - Emissiont test chamber method 
ANSI/BIFMAX7.1-2007 
StandardforFormaldehydeandTVOCEmissionsofLow-emittingOfficeFurnitureSystemsandSeating
 

III.Main Technical Parameters: 

Name & Model

XD-1VOC-A1

Operating chamber volume

1000L

Test chamber Material

2 mm thickness SUS 304 Stainless steel plate with Nanomaterials painted 

Operating      
 chamber size

Interior Dimension 1±0.02m³;W960mm×H1050mm×D1000mm

Overall dimension

W1560mm×H11600mm×D1810mm(approximately

Temperature Range

+15℃~+80℃;

Temperature fluctuations

≤±0.5℃;

Temperature uniformity

≤2℃;

Accuracy for temperature

0.1℃;

Humidity Range

5%~80%RH;

Humidity fluctuation

≤±2.5%RH;

 Accuracy for humidity

0.5%RH

Air speed in the operating chamber

Air speed of the geometric center point in the operating chamber is 0.1-0.3m /s. (Adjustable);

Air change rate

1~3times/hour (adjustable)

Purge air flow adjustment range should be above than 0.5~3 times/hour

Flowmeter accuracy

Air mass flowmeter, tolerance 0.5 grade;0~500L/min

Glass rotor flowmeter,  tolerance not less than 2.5 grade;0~3m3/h

Air Flow Display

 micro air mass flowmeter, online flow display on touch screen(adjustable) 

Airtightness

Accuracy 0.5 grade,VL < 5%×Air supply rate Vs, recovery rate:>80 %
Keep relatively positive pressure:10±5 Pa
Air pressure monitoring in the chamber, micro pressure sensor of analog quantity, pressure display on the touch screen

Conditioning period

I t takes less than 40 minutes to reach temperature of 23 °, it takes less than 70 minutes to reach humidity of 50%, and the equipment can run continuously throughout the year 

VOC concentration in air from empty chamber

Single VOC≤0.002 mg/m3,TVOC≤0.02 mg/m3

Noise

≤50db(Class A)reasonable structure, strictly control installation process, no resonance sound

Key components

        Ø Aircooling system:Danish danfoss、French Castel、American Emerson
Ø Measuring system: Finland Vaisals、JIASHANG
Ø Electrical control system:German Schneider、German Siemens、 Taiwan MEAN WELL、XIZI
Ø Pneumatic power system:German:FESTO,Japanese SMCTaiwan AirTAC

Humidity control chart

 


 

IV. Ordering Information:

Type

Name

XD-1VOC-A1

1m³ Emission Test Chamber